ELECTRONIC PROJECTS | GADGETS NEWS | LATEST TECHNOLOGY | CONSUMER ELECTRONICS
DINAMAP PRO SERIES 100-400 ALARM AND ALARM CODES
SpO,) will flash
the most recent reading and an audible alarm will be issued.
Pressing the Alarm Silence switch (causing the integral LED
to be lit) silences the audible alarm for 2 minutes, but the
alarm display reading and SILENCE LED indicator will
continue to flash at the same rate.
System Alarms
System alarms alert the operator to certain abnormal
conditions or internal system failures. Pressing the rotor
cancels the alarm information box which is displayed on the
LCD. Codes for different procedural and system alarms are
on the next page.
Failsafe Alarm
The failsafe alarm, which is the most powerful alarm of the
PRO Monitor, indicates a serious failure of the Monitor.
This alarm occurs immediately upon any failure of a self-test
and indicates system failure. When the failsafe alarm occurs,
the Monitor disables all features to ensure patient safety.
Alarm Codes
All alarm indications are accompanied by an audible signal
unless Alarm Silence is selected.
A microprocessor system failure will generate a high-pitched
audible alarm regardless of the setting of the Alarm Silence
switch.
There are three categories of alarms: patient alarms, system
alarms, and failsafe alarm.
Patient Alarms
Patient alarms include those alarms issued when the
patient’s systolic pressure, diastolic pressure, pulse rate, or
oxygen saturation is outside the set limits. Whenever one
of these conditions occurs, the associated display
(SYSTOLIC, MAP, DIASTOLIC, PULSE, or
Subscribe to:
Post Comments (Atom)
No comments:
Post a Comment